App Store Google Play YouTube X Facebook Instagram LinkedIn
Catalogue › Flammability Testing Instruments › Aviation Flammability Test Chamber — Appendix F, Part I
Aviation Flammability Test Chamber — Appendix F, Part I

Photographs have been adapted for the website; dimensions may vary.

TXL-YAN-015

Aviation Flammability Test Chamber — Appendix F, Part I

Havacılık Yanıcılık Test Kabini — Ek F Bölüm I

Appendix F, Part IVertical · horizontal · 45° · 60°
Control7-inch touchscreen PLC/HMI
Flame verificationType K thermocouple included

Flammability test chamber for the FAA 14 CFR 25 / EASA CS-25 / CAAC Appendix F, Part I Bunsen burner tests: cabin and cargo compartment materials and electrical wire and cable insulation. One chamber with four interchangeable specimen holders — vertical 12/60 s, horizontal 15 s, 45° panel 30 s and 60° wire and cable 30 s. Automatic burner advance and retraction, 7-inch touchscreen PLC/HMI, flame temperature verification with a type K thermocouple and a built-in exhaust fan.

The instrument is test apparatus; it does not carry a certificate of conformity. FAA / EASA acceptance depends on the laboratory's own procedures and the approving authority.

Model technical data

  • Test configurations: vertical 12 s / 60 s (b)(4) — TL-A016; horizontal 15 s (b)(5) — TL-A017; 45° panel 30 s (b)(6) — TL-A018; 60° electrical wire and cable 30 s (b)(7) — TL-A019.
  • Specimen holders: the chamber and the holders are separate order items; each holder can be purchased individually.
  • Evaluated results: afterflame time, burn length and flaming drips (vertical); burn rate (horizontal); flame penetration, flame time and glow time (45°).
  • Burner positioning: the operator sets the burner angle manually; the burner advances automatically to the test position and retracts when the flame application time ends.
  • Control: 7-inch touchscreen PLC/HMI for test program selection and flame application timing (12, 15, 30, 60 s); emergency stop button.
  • Gas line: methane line with regulator, pressure gauge, flowmeter and needle valve for flame adjustment; burner inlet pressure 17 ± 2 kPa.
  • Flame temperature verification: type K thermocouple included, readout on the 7-inch HMI. FAA minimums: 1550 °F (≈843 °C) for vertical, horizontal and 45° tests; 1750 °F (≈954 °C) for the 60° wire and cable test.
  • Exhaust: the built-in fan stays off during testing and clears smoke from the chamber after the test.
  • Chamber: stainless steel interior, tinted viewing window; painted exterior in RAL 7035 and RAL 7016.
  • Power supply: 230 V AC / 50 Hz or 120 V AC / 60 Hz — specify the version when ordering.

Physical specifications and utilities

Power supply230 V AC / 50 Hz or 120 V AC / 60 Hz; specify when ordering.
Current (120 V version)Approx. 3.7 A calculated; the exact rating is determined by measurement on the unit.
Gas requirementMethane (cylinder or laboratory line, supplied by the customer); burner inlet pressure 17 ± 2 kPa.
Dimensions and weightConfirmed with the order confirmation.

Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.

Product technical specifications and images may vary according to manufacturing conditions and applicable standards.

Test and configuration

Measured output / function
Afterflame time, burn length, flaming drips; burn rate; flame penetration, flame time and glow time
Parameters selected for the method
Test method and specimen holder (vertical, horizontal, 45°, 60°), flame application time, burner angle
Verification points
Flame temperature (type K thermocouple), burner inlet pressure 17 ± 2 kPa, flame application time

Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.

Control and recording specification

The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.

  • Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
  • The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
  • Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
  • Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Download the technical brochure (PDF)Rev.05 · 18.11.2026

Test standards

  • 14 CFR 25.853 Compartment interiors — Appendix F, Part I
  • 14 CFR 25.855 Cargo or baggage compartments — Appendix F, Part I
  • 14 CFR 25.1713 Electrical wire and cable insulation, (c) — 60° test
  • EASA CS-25 Large aeroplanes — Appendix F, Part I
  • CAAC CCAR-25-R4 Appendix F, Part I (b)(4)–(7)
  • AITM 2-0002 Airbus ABD0031 — vertical Bunsen burner test (A: 60 s, B: 12 s)
  • AITM 2-0003 Airbus ABD0031 — horizontal Bunsen burner test
  • AITM 2-0004 Airbus ABD0031 — 45° Bunsen burner test
  • AITM 2-0005 Airbus ABD0031 — 60° wire and cable test
  • BSS 7230 Boeing — F1–F6 Bunsen burner flammability methods
  • BSS 7324 Boeing — wire and cable flammability

Prices start from €18. Request the technical specification and a quotation for this instrument.

Request a quoteCall usPDF catalogue

More in this category

45° Flammability Tester
TXL-YAN-002

45° Flammability Tester

Horizontal Flammability Tester
TXL-YAN-003

Horizontal Flammability Tester

NFPA 701 Flame Propagation Tester
TXL-YAN-004

NFPA 701 Flame Propagation Tester

UL94 / IEC 60695 Burn Tester
TXL-YAN-005

UL94 / IEC 60695 Burn Tester

Other instruments in this group