Photographs have been adapted for the website; dimensions may vary.
Blanket Flammability Tester
Battaniye Yanma Test Cihazı
Cabinet for controlled flame contact and measurement of flame-spread time in blanket textiles. Verify flame-application control and spread timer separately.
Model technical data
- Many different test methods can be installed help with touch screen control unit and computer connection.
Physical specifications and utilities
| Electrical supply | Approx. 220–240 V AC, 50/60 Hz, single phase; build-dependent. |
|---|---|
| Power / connected load | Approx. 20–100 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 1000 × 800 × 1300 mm (W × D × H). |
| Instrument weight | Approx. 50–90 kg net. |
| Gross / shipping weight | Approx. 80–137.5 kg gross; estimated packing allowance. |
| Lab requirements | Method-specified test gas, regulator and exhaust / fume hood; gas type and flow follow the selected method and burner. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Burning behaviour of blanket specimens
- Parameters selected for the method
- Specimen orientation, flame application and conditioning
- Verification points
- Burner/gas flow, geometry and timing
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ASTM D4151-92 (2001) · 6.1–6.1.1.2
- Control flame contact in the blanket cabinet to 1 ±0.05 seconds. The flame-spread counter reads to 0.01 seconds; counter resolution is distinct from flame-contact timing accuracy. Verify both timing functions separately.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.
Request a quoteCall usPDF catalogue