Climatic Conditioning Chamber
İklim Koşullandırma Kabini
Conditioning chamber at standard conditions (20°C±2 / 65%RH±4). Instantly and accurately detects moisture and temperature controls, which are critically important for sensitive production environments. Offers portable and practical use.
Model technical data
- The interior is made from stainless steel.
Physical specifications and utilities
| Electrical supply | Approx. 220–240 V AC, 50/60 Hz, single phase; build-dependent. |
|---|---|
| Power / connected load | Approx. 1500–3500 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 900 × 850 × 1800 mm (W × D × H). |
| Instrument weight | Approx. 130–250 kg net. |
| Gross / shipping weight | Approx. 168–337.5 kg gross; estimated packing allowance. |
| Lab requirements | Ventilated installation; purified water and drain for humidification options. Separate conditioning plant depends on order scope. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Specimen conditioning at controlled temperature and relative humidity
- Parameters selected for the method
- Target temperature, relative humidity, load and conditioning time
- Verification points
- Temperature-humidity mapping and temporal stability
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ASTM D1776-98e1 · 6.1, Table 1
- Table: general textiles at 21 ± 1 °C and 65 ±2% RH; nonwovens at 23 ± 1 °C and 50 ±2% RH. Record chamber temperature and humidity; keep these conditions separate from the ISO 139 recipe.
ISO 139:2005 + Amd 1:2011 · 3-5, Annex A
- Standard atmosphere 20 °C / 65% RH; agreed alternatives 23 °C / 50% RH or 27 °C / 65% RH, tolerances ±2 °C and ±4% RH. Recommended monitor resolution 0.1 °C and 0.1% RH or better; measurement uncertainty ±0.5 °C and ±2% RH or better. Assess monitoring sensors separately from control sensors.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.