Catalogue › Washing Fastness & Lab Dyeing Machines › Washing Fastness Tester (Gyrowash)
Washing Fastness Tester (Gyrowash)

Photographs have been adapted for the website; dimensions may vary.

TXL-BOY-001

Washing Fastness Tester (Gyrowash)

Yıkama Haslığı Test Cihazı

Rotary tumbling washing fastness. ISO 105-C / AATCC.

Model technical data

  • Number of test vessels: 6 standard; 8/12/16/20/24 tube options available.
  • Test vessel volume: 550 ml.
  • Rotation speed: 40 ± 2 rpm.
  • Operating temperature: ambient – 98 °C.
  • Temperature accuracy: ± 1 °C.
  • Programmable memory-based control system with 3.5" touch screen.
  • Colour fastness to washing, dry cleaning and chlorinated water.

Physical specifications and utilities

Electrical supplyApprox. 220–240 V AC, 50/60 Hz, single phase; build-dependent.
Power / connected loadApprox. 2500–5500 W connected load for preliminary planning.
Overall instrument dimensionsApprox. 850 × 800 × 1050 mm (W × D × H).
Instrument weightApprox. 90–150 kg net.
Gross / shipping weightApprox. 124–212.5 kg gross; estimated packing allowance.
Lab requirementsWater supply and drain; hot water, water quality and plumbing capacity depend on order configuration.

Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.

Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.

Product technical specifications and images may vary according to manufacturing conditions and applicable standards.

Test and configuration

Measured output / function
Colour change and staining of adjacent fabric
Parameters selected for the method
Vessel volume, temperature, time, detergent and balls
Verification points
Bath temperature, rotational speed and timer

Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.

Test parameters for the reviewed method

Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.

ISO 105-C06 (earlier edition) · 4.1-4.2

  • ISO vessel arrangement: 550 ± 50 ml; diameter 75 ± 5 mm and height 125 ± 10 mm; rotation 40 ± 2 r/min; solution temperature within ±2 °C of the selected value.

ISO 105-C08:2010 · 5.1-5.2

  • Wash vessel 550 ± 50 mL; diameter 75 ± 5 mm, height 125 ± 10 mm; rotation 40 ± 2 r/min. Solution within ±2 °C of the selected test temperature; weighing accuracy 0.01 g.

TS EN ISO 105-C09:2003 · 5.1-5.3

  • Recipe: vessel rotation 40 ± 2 r/min and solution-temperature deviation no greater than ±2 °C. Preparation mixer at least 1000 r/min and balance precision ±0.01 g; auxiliary preparation equipment is selected separately.

TS EN ISO 105-D01:2011 · 6.1, 8

  • Dry-cleaning recipe: 550 ± 50 mL vessel with solvent-resistant seals, 40 ± 2 r/min and 30 ± 2 °C. Solvent and required fume hood are assessed separately from the aqueous-washing configuration.

AATCC 61-2009 · 5.1.1-5.1.5

  • Rotation 40 ± 2 r/min; vessels differ: 500 mL (75 × 125 mm) for 1A and 1200 mL (90 × 200 mm) for 1B/2A/3A/4A/5A. Steel-ball diameter 6 mm; vessel and abrasive set are recorded for the selected test number.

TS EN ISO 105-C10:2011 · 4.1-4.4

  • Vessel 550 ± 50 mL, diameter 75 ± 5 mm, height 125 ± 10 mm; rotation 40 ± 2 r/min and solution within ±2 °C of the selected temperature. Approximately 6 mm steel balls and soap/alkali recipe recorded for the selected test condition.

TS EN ISO 105-E03:2010 · 4.1

  • Chlorinated-water fastness arrangement: glass or stainless 550 ± 50 mL vessel, 40 ± 2 r/min and ±2 °C solution-temperature control. Active-chlorine level and solution pH are tracked in separate recipe fields.

M&S C37:2006-02 · Apparatus

  • Chlorinated-water fastness: 500 ± 50 mL vessel, chlorine-resistant seal, 27 ± 2 °C and 40 ± 2 r/min. This edition excludes infrared-heated wash wheels.

M&S C10A — June 2007 · Apparatus 1–3; Specimen; Wash liquor; Method 1–8; Note5; pp.1–4

  • M&S C10A June 2007: vessel capacity 550 ± 50 mL, rotation 40 ± 2 rpm; infrared-heated wash wheels are not accepted in this edition. These historical method requirements do not automatically apply to sample-dyeing machines. The auxiliary stirrer must sustain about 1000 rpm for at least 10 min; balance accuracy at least ±0.01 g.
  • Fabric specimen 100 × 40 mm; liquor ratio 100:1 (100 mL solution per 1 g fabric). Start at 27 ± 2 °C and ramp at 1.5 ± 0.5 °C/min. A 40 °C care label uses a 50 °C test; a 50 °C label uses 60 °C. Start the 30 min hold on reaching the target. These method volumes/specimen masses are not machine capacity or net weight.
  • The method assesses oxidative-bleach colour change; no multifibre adjacent fabric or staining assessment. Prepare fresh solution for immediate use and adjust TAED mass to documented activity. Drying limit 60 °C. Current M&S approval and delivered ramp accuracy require separate verification.

Control and recording specification

The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.

  • Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
  • The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
  • Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
  • Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
  • The C10A record links care-label/test temperatures, measured ramp and a 30 min hold starting at target temperature. Trace specimen mass/liquor volume and reagent activity; do not add optical-brightener effects or multifibre staining scores to this result. These are method requirements; delivered modules are specified separately.
Download the technical brochure (PDF)Revision: 27 October 2026

Test standards

Prices start from €18. Request the technical specification and a quotation for this instrument.

Request a quoteCall usPDF catalogue

More in this category

Infrared Sample Dyeing Machine
TXL-BOY-002

Infrared Sample Dyeing Machine

Sample Dyeing Machine (Oil-Bath)
TXL-BOY-003

Sample Dyeing Machine (Oil-Bath)

Package Dyeing Machine (Lab)
TXL-BOY-004

Package Dyeing Machine (Lab)

Drum-Type Piece Fabric Dyeing Machine
TXL-BOY-005

Drum-Type Piece Fabric Dyeing Machine