M&S C10A testing instrument
M&S C10A June 2007 was reviewed: 550 ± 50 mL vessels, 40 ± 2 rpm; infrared heating is not accepted in this edition. Verify the 100:1 liquor ratio and 30 min hold starting at target temperature for TXL-BOY-001. The method does not include multifibre staining scores. Current M&S approval is not verified.
About the standard
| Code | M&S C10A |
|---|---|
| Published by | Marks & Spencer |
| Source | scribd.com |
Other standards tested on the same instrument
AATCC 61 · ISO 105-C06 · ISO 105-C08 · ISO 105-C09 · ISO 105-C10 · ISO 105-D01 · ISO 105-E03 · M&S C04 · M&S C05 · M&S C12A · M&S C37
FAQ
Which instrument is used for M&S C10A testing?
M&S C10A June 2007 was reviewed: 550 ± 50 mL vessels, 40 ± 2 rpm; infrared heating is not accepted in this edition. Verify the 100:1 liquor ratio and 30 min hold starting at target temperature for TXL-BOY-001. The method does not include multifibre staining scores. Current M&S approval is not verified.