Photographs have been adapted for the website; dimensions may vary.
Carpet Flammability Tester
Halı Yanmazlık Test Cihazı
A test chamber for evaluating the effects of a small ignition source on carpets and floor coverings. Methenamine tablet or hot metal nut testing is performed with the appropriate fixture.
Model technical data
- Assesses the response of textile floor coverings to a small ignition source.
- A method-specific setup is selected for methenamine tablet or hot metal nut testing.
- The test chamber is made of non-combustible, heat-resistant panels.
- Counterbalanced sliding access panel and upper observation mirror.
Physical specifications and utilities
| Electrical supply | Manual / passive base apparatus; no mains supply assumed. |
|---|---|
| Power / connected load | Approx. 0 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 500 × 500 × 500 mm (W × D × H). |
| Instrument weight | Approx. 15–30 kg net. |
| Gross / shipping weight | Approx. 26.5–47.5 kg gross; estimated packing allowance. |
| Lab requirements | Suitable ventilation / process exhaust; external fan and duct loads are excluded from this preliminary estimate. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Carpet damage after a small ignition source
- Parameters selected for the method
- Separate kit for methenamine tablet or hot-nut method
- Verification points
- Burner/gas flow, geometry and timing
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ISO 6925:1982 / SIST ISO 6925:1998 · 4.1-4.9
- Carpet tablet test: open-top chamber, internal size 300 × 300 × 300 mm; metal plate 230 × 230 mm, thickness 6.5 ± 0.5 mm, opening Ø205 mm. Methenamine tablet 150 ± 5 mg, Ø6 mm; chamber dimensions are not overall instrument dimensions.
BS 6307:1982 · 4.1–4.2, 4.4, 4.8, 6.2
- Method: open-top test box internally 300 × 300 × 300 mm; plate 230 × 230 mm, thickness 6.5 ± 0.5 mm, opening 205 mm. Separate viewing hood about 2 m³; extraction off during test. Dry-conditioning option requires a separate oven at 105 ± 2 °C.
BS 4790:1987 + amendments 6444/7376 — selected legible clauses · Appendix A; 9.3.2–9.3.4; 11(e)–(f)
- Annex A: open-top box internally 380 × 380 × 380 mm; draught-free surroundings. Measure affected-area radius from nut application centre; report use face and observable underside separately, greatest radius to nearest 5 mm. Nut and heating tolerances remain outside this legible-clause review.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.
Request a quoteCall usPDF catalogue