External Balance Water Vapour Transmission Rate Tester (WVTR-08E)

Photographs have been adapted for the website; dimensions may vary.

TXL-MBR-006

External Balance Water Vapour Transmission Rate Tester (WVTR-08E)

Harici Terazili Su Buharı Geçiş Hızı Test Cihazı (WVTR-08E)

Gravimetric water vapour transmission rate tester with eight test positions and an external analytical balance. The operator only weighs the cups; climate control, cup indexing, data transfer and reporting are handled by the system. To ASTM E96/E96M.

Model technical data

  • Reference standard: ASTM E96/E96M-24a — gravimetric water vapour transmission rate.
  • Test positions: 8; suitable for a 6 specimen + 2 blank arrangement.
  • Cup / test area: hard-anodised aluminium cup; Ø76 mm clear opening; 45.36 cm².
  • Specimen: approx. Ø90 mm cut; standard 0-3 mm thickness.
  • Temperature: 20-50 °C; stability ±0.5 °C.
  • Relative humidity: 40-95% RH; stability ±2% RH.
  • Air velocity: 0.05-1.00 m/s, closed-loop adjustable horizontal airflow.
  • Rotary platform: 0.5-1 rpm continuous rotation with automatic cup indexing.
  • External analytical balance: 220 g capacity; 0.1 mg readability; internal calibration; RS-232 and USB.
  • Approx. dimensions and power: 1000 × 900 × 1650 mm; 230 V AC, 50 Hz; installed power approx. 2.5 kW.

Physical specifications and utilities

Electrical supplyApprox. dimensions and power: 1000 × 900 × 1650 mm; 230 V AC, 50 Hz; installed power approx. 2.5 kW.
Power / connected loadApprox. dimensions and power: 1000 × 900 × 1650 mm; 230 V AC, 50 Hz; installed power approx. 2.5 kW.
Overall instrument dimensionsApprox. dimensions and power: 1000 × 900 × 1650 mm; 230 V AC, 50 Hz; installed power approx. 2.5 kW.
Instrument weightApprox. 120–220 kg net.
Gross / shipping weightApprox. 157–300 kg gross; estimated packing allowance.
InterfacesExternal analytical balance: 220 g capacity; 0.1 mg readability; internal calibration; RS-232 and USB.
Lab requirementsVentilated installation; purified water and drain for humidification options. Separate conditioning plant depends on order scope.

Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.

Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.

Product technical specifications and images may vary according to manufacturing conditions and applicable standards.

Test and configuration

Measured output / function
Water vapour transmission; g/(m²·day) or method-specific index
Parameters selected for the method
Cup area, temperature, humidity, sealing and weighing interval
Verification points
Balance, temperature-humidity, cup area and blank test

Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.

Test parameters for the reviewed method

Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.

ASTM E96/E96M-16 · 6.1-6.3, 11-13

  • Cup method: opening at least 3000 mm²; air velocity over the specimen 0.02-0.3 m/s. Balance sensitive to less than 1% of the steady-period mass change. Calculate WVTR = mass-time slope / area from at least six suitable steady points; equivalence to edition 24a is not assumed.

ISO 12572:2001 · 5

  • Cup method: chamber control within ±0.5 K and ±3% RH of target, air speed 0.02–0.30 m/s. Prefer 0.001 g balance resolution, selected for assembly mass. This is the 2001 apparatus clause; equivalence to the 2016 edition and 2024 amendment is not assumed.

Control and recording specification

The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.

  • Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
  • For WVTR, mass-time slope / effective area; cup method, time unit and temperature-humidity conditions are stored with the result.
  • Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
  • Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Download the technical brochure (PDF)Revision: 27 October 2026

Test standards

Prices start from €18. Request the technical specification and a quotation for this instrument.

Request a quoteCall usPDF catalogue

More in this category

CryoBend — Cold Bending Tester
TXL-MBR-001

CryoBend — Cold Bending Tester

Water Vapour Transmission Rate Tester (WVTR-01A)
TXL-MBR-002

Water Vapour Transmission Rate Tester (WVTR-01A)

EN 1928 Waterproofing Impermeability Tester
TXL-MBR-003

EN 1928 Waterproofing Impermeability Tester

EN 12310-1 Nail-Shank Tear Test Fixture
TXL-MBR-004

EN 12310-1 Nail-Shank Tear Test Fixture