Photographs have been adapted for the website; dimensions may vary.
EN 1928 Waterproofing Impermeability Tester
EN 1928 Su Yalıtım Test Cihazı
Reliable water impermeability testing for waterproofing membranes. Compact and precise solution performing Method A and Method B under EN 1928 on a single platform. 0–200 kPa pressure control, 24-hour stable test capacity, stainless steel housing, digital control panel, interchangeable fixture system. Test duration: 24±1 hours. Ideal for roof membranes, bituminous systems, PVC/TPO membranes.
Model technical data
- Method A and Method B in a single device.
- Pressure range: 0-200 kPa.
- Test duration: 24 ± 1 hours.
- Specimen diameter: Ø150 mm (Method A), Ø130 mm (Method B).
- Interchangeable fixture system.
- Stainless steel body.
- Digital control panel with display.
- Power supply: 220 V ± 10%, 50/60 Hz.
Physical specifications and utilities
| Electrical supply | Power supply: 220 V ± 10%, 50/60 Hz. |
|---|---|
| Power / connected load | Approx. 100–300 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 900 × 750 × 650 mm (W × D × H). |
| Instrument weight | Approx. 60–110 kg net. |
| Gross / shipping weight | Approx. 91–162.5 kg gross; estimated packing allowance. |
| Lab requirements | Suitable water supply / drain; dry filtered compressed air for pneumatic fixture option. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Watertightness under specified water pressure
- Parameters selected for the method
- Method, pressure, duration, specimen count and support
- Verification points
- Pressure, timing and chamber sealing
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
TS EN 1928:2007 · 5, 7.2, 8
- Methods A and B require separate fixtures. In this edition the Method A specimen diameter is 200 ± 2 mm; test temperature 23 ± 1 °C and pressure hold 24 ± 1 h. Pressure is recorded according to the relevant product specification.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.
Request a quoteCall usPDF catalogue