Photographs have been adapted for the website; dimensions may vary.
Digital Yarn Count Balance
Dijital İplik Numarası Terazisi
Yarn count system consisting of an electronic balance and dedicated software. Calculates yarn, sliver and roving counts in Tex, Denier, Nm, Ne and other systems, and fabric mass per unit area in g/m²; gives totals, average, CV and standard deviation.
Model technical data
- Electronic balance: 300 g capacity, 0.001 g readability (0.1 mg balance option).
- Yarn, sliver and roving count: Tex, Denier, Nm, Ne and other systems; fabric mass per unit area (g/m²).
- Statistics: total, arithmetic average, CV %, standard deviation, maximum and minimum, range.
- New recipes are created by saving the number of tests and the sample properties.
- Touch screen and built-in 58 mm printer.
- Length options 3 yd, 7 yd, 120 yd, 840 yd, 2 m, 10 m, 100 m and 500 m.
- Connects to the balance with a data cable; USB, cables and security data key included; remote control over the internet.
- Used together with the yarn wrap reel, which can be supplied on request.
Physical specifications and utilities
| Electrical supply | Approx. low-voltage adapter with 100–240 V AC, 50/60 Hz input; output voltage is build-dependent. |
|---|---|
| Power / connected load | Approx. 10–25 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 350 × 350 × 300 mm (W × D × H). |
| Instrument weight | Approx. 3–7 kg net. |
| Gross / shipping weight | Approx. 5.3–10.8 kg gross; estimated packing allowance. |
| Interfaces | Connects to the balance with a data cable; USB, cables and security data key included; remote control over the internet. |
| Lab requirements | Level, stable working surface; specimen conditioning follows the selected method. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Linear density or yarn count (tex, den, Nm, Ne)
- Parameters selected for the method
- Skein length, mass and selected count system
- Verification points
- Reel circumference, turn count and balance
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ISO 2060:1994 · 5.1 / Ek A
- Recommended reel circumference: 1.000 m ± 2.5 mm.
- Pretension with positive feed: 0.5 ± 0.1 cN/tex.
ISO 3801:1977 · 4.4-4.5
- Small-specimen method: 10 × 10 cm template or 100 cm² circular cutter with 1% area accuracy; balance capable of weighing to 0.001 g. Method and specimen area are stored with the mass-per-area calculation.
ASTM D1907-97 · 6.1
- For the skein method, recommended reel perimeter 1.000 m or 1.500 yd with ±0.25% tolerance. Record turns, actual perimeter, tension and weighing condition together; use actual wound length for linear-density calculation.
ASTM D2646-96 · 1, 14.1–14.2
- This link concerns the backing-fabric mass-per-area subtest: D3776 for woven and D3887 for knitted fabric. Balance software retains area and g/m²; the reference does not mean the balance performs every mechanical test in this standard.
ASTM D3776-96 · 1.2, 5.1–5.2
- Distinguish full-piece, full-width, small-swatch and narrow-fabric options. Balance capacity/sensitivity must support weighing within ±0.1% of specimen mass. With a cutter, retain actual cut area, mass and calculated g/m².
ISO 9073-1:2023 · 5, 7-9
- Nonwoven mass per area: specimen area at least 50,000 mm², at least three specimens; weighing accuracy ±0.1% of mass. The balance is the weighing component of a method also requiring cutting and area measurement.
EN 12127:1997 / SIST EN 12127:1999 · 6, 8.2-8.4, 9-10
- Small-fabric-specimen mass per area: cutter area at least 100 cm² with ±1% area accuracy, ±1 mg balance and at least five specimens. Calculate using actual specimen area; distinguish standard-atmosphere mass from oven-dry mass.
ISO 3374:2000 · 5.1-5.4
- Reinforcement mat area 1000 cm², reinforcement fabric 100 cm²; area error at most 1%. Fabric ≥200 g/m² requires 1 mg balance resolution; <200 g/m² requires 0.1 mg. Use the 0.1 mg balance option for lightweight fabric.
BS 2471:1978 + Amd.9955:1998 — historical selected clauses · Foreword; 3.3; 4.1–4.4; methods 1–4 only
- 1978/1998 edition, methods 1–4: balance accuracy ±0.2% of measured mass; calibrated steel rule 3 m, or ≥2 m if unavailable. Flat table wider than fabric, length ≥4 m. Method 5 was withdrawn in this edition; do not use it as a small-cut GSM method. Verify the 2005 edition separately.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- The need for integrated software depends on the method and selected hardware; manual instruments use an external test record.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- tex = 1000 × mass(g) / length(m); denier = 9000 × mass(g) / length(m).
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.
Request a quoteCall usPDF catalogue