Physical specifications and utilities
| Electrical supply | Approx. 220–240 V AC, 50/60 Hz, single phase; build-dependent. |
|---|---|
| Power / connected load | Approx. 200–500 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 750 × 650 × 1200 mm (W × D × H). |
| Instrument weight | Approx. 90–150 kg net. |
| Gross / shipping weight | Approx. 124–212.5 kg gross; estimated packing allowance. |
| Lab requirements | Level, stable working surface; specimen conditioning follows the selected method. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Thickness loss after repeated impacts
- Parameters selected for the method
- Impact count, loading element and specimen position
- Verification points
- Impact counter, loading element and thickness measurement
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ISO 2094:1999 · 5.1.1-5.2
- Dynamic carpet loading: impact mass 1279 ± 13 g, drop height 63.5 ± 0.5 mm and interval 4.3 ± 0.3 s. Configure the two feet and traversing table to the method; count impacts. Measure thickness at 2.0 kPa with 0.1 mm accuracy.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- Thickness loss and recovery are calculated using initial thickness and method-specific times.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
- Record schema: 50/100/200/1000 impact stages, initial/final thickness and thickness loss, following the ISO 2094 recipe.
Test standards