UV Weathering / Ageing Tester
UV Yaşlandırma Test Cihazı
Uses fluorescent UV lamps, temperature and moisture cycles for accelerated material weathering. Condensation and water spray steps simulate outdoor exposure.
Model technical data
- Principle: fluorescent UV, temperature and moisture cycles.
- Lamp arrangement: 4 per side, 8 fluorescent UV lamps in total.
- Lamp options: UVA-340, UVB-313 or UVA-351, selected with the method.
- UV irradiance: closed-loop control with sensor feedback.
- Operator interface: 7-inch touchscreen PLC/HMI.
- Moisture stages: condensation and water spray.
- Data interface: Ethernet; PC software and printer are optional.
- Access from both sides; lamp family and irradiance calibration are specified together.
Physical specifications and utilities
| Electrical supply | Approx. 220–240 V AC, 50/60 Hz, single phase; build-dependent. |
|---|---|
| Power / connected load | Approx. 1500–2500 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 1400 × 600 × 1450 mm (W × D × H). |
| Instrument weight | Approx. 120–180 kg net. |
| Gross / shipping weight | Approx. 157–250 kg gross; estimated packing allowance. |
| Interfaces | Data interface: Ethernet; PC software and printer are optional. |
| Lab requirements | Ventilation; suitable purified water and drain for condensation / spray options. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Colour, surface and property changes after UV exposure
- Parameters selected for the method
- Lamp family, irradiance, temperature, time and moisture cycle
- Verification points
- Irradiance sensor, specimen temperature and cycle record
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ISO 4892-3:2013 · 4.1.1
- UVA-340, UVA-351 and UVB-313 lamps have different spectral distributions, with peaks around 343, 353 and 313 nm respectively. Record lamp type, irradiance measurement band and cycle together; recipes are not assumed equivalent.
ISO 16474-3:2013 + corrigendum 2015 · 5.1.1
- For paint/varnish UV exposure, UVA-340 (1A), UVA-351 (1B) and UVB-313 (2) sources are selected separately. Measured irradiance, lamp type and exposure cycle must remain traceable in the report.
EN 927-6:2018 · 5, 7.3.1-7.3.3
- Exterior-wood-coating cycle: 24 h condensation at 45 ± 3 °C; then 48 repetitions of 2.5 h UV (60 ± 3 °C; 0.89 W/(m²·nm) at 340 nm) and 0.5 h spray with UV off. Repeat the weekly cycle 12 times, totalling 2016 h.
ASTM G154-06 · 4.2, 6.1-6.1.3
- Fluorescent-UV exposure records lamp type, irradiance, moisture step, temperature and duration separately. UVA-340, UVA-351 and UVB-313 spectra differ; monitor specimen-plane irradiance with a radiometer. No universal cycle is specified.
ASTM D4329-99 · 5.1–5.3.3, 6.5, 6.9
- 1999 edition: G151/G154 fluorescent UV apparatus, UVA-340 unless specified otherwise. Installation ambient 18–27 °C; wall/equipment clearance ≥300 mm; remove heat/moisture by ventilation. At least three replicates; below 90% of peak irradiance, use equal-exposure or position compensation procedures.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- Irradiance, temperature and moisture stages are time-aligned; spectral band and integration interval are stated when calculating dose.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.
Request a quoteCall usPDF catalogue