Fiber Bundle Strength Tester
Elyaf Tutam Mukavemeti Test Cihazı
Measurement of fiber bundle strength properties. Strengthens quality control processes by measuring the structural properties of fibers in detail at the micro level. Offers rapid calibration capabilities.
Model technical data
- Used to analyse breaking load, tensile resistance and average fibre breaking load.
Physical specifications and utilities
| Electrical supply | Manual / passive base apparatus; no mains supply assumed. |
|---|---|
| Power / connected load | Approx. 0 W connected load for preliminary planning. |
| Overall instrument dimensions | Approx. 400 × 300 × 450 mm (W × D × H). |
| Instrument weight | Approx. 12–25 kg net. |
| Gross / shipping weight | Approx. 23.2–41.2 kg gross; estimated packing allowance. |
| Lab requirements | Level, stable working surface; specimen conditioning follows the selected method. |
Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.
Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.
Product technical specifications and images may vary according to manufacturing conditions and applicable standards.
Test and configuration
- Measured output / function
- Fibre-bundle breaking force and tenacity
- Parameters selected for the method
- Bundle preparation, jaw spacing and bundle mass
- Verification points
- Force system and specimen weighing
Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.
Test parameters for the reviewed method
Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.
ASTM D1445-95 · 6.2–6.3
- Select zero-gauge and 3.2 mm gauge procedures separately; balance sensitivity is 0.01 mg. Total clamp thickness is 11.81 ±0.02 mm and spacer thickness 3.2 ±0.02 mm. Record gauge, bundle mass and clamp arrangement with results.
Control and recording specification
The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.
- Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
- The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
- Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
- Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
Test standards
Prices start from €18. Request the technical specification and a quotation for this instrument.