M&S C9 October 2004 and its two equipment appendices were read. Specimen at least 45 × 20 mm; distinguish specimen effective humidity from chamber RH and black-panel from black-standard sensors. The Ci3000+ appendix requirement of 42 W/m² over 300–400 nm is not copied automatically to another setup. Verify TXL-KMS-031/033 hardware and current M&S approval separately.

Light Fastness Tester (Xenon Arc)

Light Fastness Tester (8 Station / 24 Specimen)
About the standard
| Code | M&S C9 |
|---|---|
| Published by | Marks & Spencer |
| Source | scribd.com |
Other standards tested on the same instrument
ISO 105-B02 · ISO 105-B04 · ISO 105-B07 · ISO 4892-1 · ASTM D6695 · ASTM G151 · ASTM G155 · EN 15187 · AATCC 16 · AATCC 169 · M&S C9A · NF G07-012-2
FAQ
Which instrument is used for M&S C9 testing?
M&S C9 October 2004 and its two equipment appendices were read. Specimen at least 45 × 20 mm; distinguish specimen effective humidity from chamber RH and black-panel from black-standard sensors. The Ci3000+ appendix requirement of 42 W/m² over 300–400 nm is not copied automatically to another setup. Verify TXL-KMS-031/033 hardware and current M&S approval separately.