Catalogue › Fibre Testing Instruments › Rapid Oil Extraction Apparatus
Rapid Oil Extraction Apparatus

Photographs have been adapted for the website; dimensions may vary.

TXL-ELY-009

Rapid Oil Extraction Apparatus

Hızlı Yağ Ekstraksiyon Cihazı

Determines the oil or finish content of wool or synthetic samples by solvent percolation and evaporation within 15 minutes. Temperature range is 60–150 °C; sample weight is 2 g per test.

Model technical data

  • Determines the oil or finish content of wool or synthetic samples by solvent percolation and evaporation within 15 minutes.
  • Temperature range is 60-150 °C and sample weight is 2 g per test.

Physical specifications and utilities

Electrical supplyApprox. 220–240 V AC, 50/60 Hz, single phase; build-dependent.
Power / connected loadApprox. 1000–2000 W connected load for preliminary planning.
Overall instrument dimensionsApprox. 700 × 500 × 700 mm (W × D × H).
Instrument weightApprox. 25–50 kg net.
Gross / shipping weightApprox. 37.5–72.5 kg gross; estimated packing allowance.
Lab requirementsSuitable fume hood, cooling water / recirculator and waste collection for solvent work; solvent follows the method.

Dimension order and approximate values are stated in each row. Packaging data are separate from instrument dimensions and net mass. Approx. rows are preliminary engineering estimates based on comparable equipment and function/capacity; they are not verified manufacturing measurements or standard requirements. Installation and shipment are finalised against the approved drawing and nameplate. Estimate scope: base product / one unit; external computer, bench and auxiliary plant excluded.

Specifications, dimensions, and power requirements are approximate and subject to change without prior notice for continuous design improvements.

Product technical specifications and images may vary according to manufacturing conditions and applicable standards.

Test and configuration

Measured output / function
Solvent-extractable oil or finish (%)
Parameters selected for the method
Solvent, sample mass, temperature and duration
Verification points
Balance, temperature and solvent blank

Listed methods are application references. Suitability is established by assessing the selected hardware, current method and calibration together.

Test parameters for the reviewed method

Values belong to the stated source edition. Test recipe, fixture and delivered instrument configuration are selected together.

GB/T 6504:2017 · 5.1.4-5.1.6, Annex C.4-C.6

  • Fibre oil-content Annex C rapid pressure extraction: approximately 2 g specimen, balance readability 0.1 mg and temperature control ±1 °C; initial solvent soak 10 min. Classical Method A requires at least 2 h reflux; the instrument's approximate 15 min cycle is not every method's duration.

GB/T 6977:2008 · 1–7.5 read; ethanol application: 3.5, 4.1, 5(a–d), 6(a), 7.1–7.2, 7.5

  • Scoured-wool ethanol extractives: three separate 5 g specimens; pre-dry at 105 ± 2 °C for 2 h, then to constant mass. Soxhlet with a constant-temperature water bath; 95% ethanol by mass, at least 20 reflux cycles at approximately 6–7 cycles/h. Dry the extract flask at 105 ± 2 °C to constant mass.
  • Auxiliary analytical balance: 200 g capacity, 0.001 g graduation; test weighings to 0.001 g. Constant mass: the difference between consecutive weighings does not exceed 0.05% of the later mass. The balance and drying oven are separate laboratory equipment; these are not instrument weight or total electrical load.
  • Configuration limit: the model’s 2 g / approximately 15 min rapid extraction data are not equivalent to this Soxhlet method. A suitable 5 g Soxhlet configuration, duration and auxiliary equipment must be verified for GB/T 6977. The standard’s ash and alkali-insoluble tests require separate apparatus.

Control and recording specification

The following are method-specific software/recording requirements; delivered software modules are defined by the order scope.

  • Method code and edition, specimen ID, operator and selected test parameters are linked to the record.
  • The calculation and endpoint are defined by the selected method edition; instrument signals and operator assessments are kept distinct.
  • Raw measurements, units, valid/invalid test flags and the reported result remain traceable.
  • Calculations are checked against known inputs and an independent calculation; sensor channels are checked against calibration values.
  • GB/T 6977:2008 ethanol recipe: E = 100 × (m2 − m1) / m3; m2 is the flask after extraction, m1 the empty flask and m3 dry wool mass. Retain three replicate results and constant-mass weighings; calculate to three decimals and round the mean to two under GB/T 8170. Ash calculation is outside this module’s scope.
Download the technical brochure (PDF)Revision: 27 October 2026

Test standards

Prices start from €18. Request the technical specification and a quotation for this instrument.

Request a quoteCall usPDF catalogue