LAB PRACTICE

Calibration, verification and intermediate checks: what a textile lab actually has to do between certificates

October 28, 2026 · 5 min read

Most textile laboratories have a calibration folder. Fewer have a system that tells them, on any given Tuesday, whether the tensile tester is still reading correctly. The difference between the two is the subject of this article.

The reference document is ISO/IEC 17025:2017, the general requirements for the competence of testing and calibration laboratories. You do not need to be accredited to use its logic; it is simply the most widely recognised description of how to keep measurement equipment under control.

Four words that are used as if they meant the same thing

TermWhat it isWho usually does it
CalibrationComparing the instrument's reading with a reference of known value, and stating the difference and its uncertaintyA competent calibration laboratory
AdjustmentChanging the instrument so that it reads closer to the referenceThe calibrator or the manufacturer
VerificationDeciding whether the instrument meets a specified requirement — for example a tolerance in a test standardThe laboratory, using calibration results
Intermediate checkA simpler check, between calibrations, that the instrument has not driftedThe laboratory itself

A certificate that says "calibrated" without telling you whether the instrument meets the tolerance your test method requires is only half the job. The verification — comparing the result against the requirement — is yours.

What ISO/IEC 17025 asks for

The equipment clauses of the standard are short and worth reading in full. In summary:

Intermediate checks also appear in clause 7.7.1 among the ways a laboratory monitors the validity of its results.

A textile instrument is several instruments

A tensile tester, a wash-fastness tester or a Martindale is not calibrated as a single object. It is a set of measured quantities, each with its own reference:

InstrumentQuantities that affect the result
Tensile / tear testerForce, extension or crosshead travel, speed
Wash-fastness tester, washing machineBath temperature, time, rotation, liquor volume
Martindale, crockmeterMass of loading pieces, cycle count, stroke
Conditioning room or cabinetTemperature and relative humidity
BalancesMass
Flammability cabinetsTime, flame height, gas flow
Light cabinetLight source condition and hours in use

For each quantity, write down three things: the tolerance the test method requires, how it is calibrated, and how it is checked in between. For tensile testers, the force-measuring system is normally verified to ISO 7500-1, which classifies the machine by its accuracy; check that the class on the certificate is good enough for the methods you run.

Designing intermediate checks that people will actually do

An intermediate check is useful only if it is quick enough to be done when it should be. Some practical patterns:

Set an acceptance limit for every check before you start, and write down what happens when it is exceeded.

Setting and changing intervals

There is no universal calibration interval. Start with the manufacturer's recommendation, the requirements of the standards you run and your customers' expectations. Then use your own data: an instrument whose intermediate checks never move can justify a longer interval; one that drifts needs a shorter one. ISO/IEC 17025 explicitly expects the programme to be reviewed and adjusted.

Reading a calibration certificate

Before you file it, check:

  1. Range — were the calibration points inside the range you actually use?
  2. As found / as left — was the instrument adjusted? If so, what did it read before adjustment? That is what your results were based on.
  3. Uncertainty — is it small enough compared with the tolerance you need?
  4. Traceability — does the certificate state the reference and its traceability?
  5. Conformity — has someone in your laboratory compared the results with the method tolerance and signed the decision?

When something is out

If a calibration or intermediate check shows the instrument outside its limits, stop using it, label it, and go back through the results produced since the last good check. Decide whether they are affected and whether customers need to be told. This is uncomfortable, and it is exactly why frequent, simple checks are worth more than an annual certificate alone.

Grey scales and other visual references deserve the same treatment: they fade and get handled, and they should be inspected and replaced on a plan.

We build our instruments in our own workshop in Istanbul and supply the technical data you need to define calibration points and check procedures for each of them.

Cenk YILMAZ
TLI-TexLabInstruments