October 28, 2026 · 4 min read

Every fastness test in the ISO 105 series — washing, rubbing, perspiration, light — ends the same way: someone holds a grey scale next to a specimen and writes a number between 1 and 5.

That number is where most disputes start. The test itself was done correctly; the reading was not. The good news is that the reading has rules too, and most of them fit on one page.

Two scales for two questions

ISO 105-A02ISO 105-A03
QuestionHow much did the specimen change colour?How much colour did the adjacent fabric pick up?
PairsGrey on grey; the second chip gets lighterWhite on white; the second chip gets darker
Typical useColour change after washing, light, perspirationStaining of multifibre or rubbing cloth

Using the change-in-colour scale to grade staining, or the reverse, is not a small slip. The steps are spaced differently.

What a "grade" actually is

Each step of the scale is a pair of chips with a specified colour difference in CIELAB units. The 9-step versions add the half grades.

GradeA02 colour change (CIELAB ΔE)A03 staining (CIELAB ΔE)
500
4–50.82.2
41.74.3
3–42.56.0
33.48.5
2–34.812.0
26.816.9
1–29.624.0
113.634.1

Values from ISO 105-A02:1993 and ISO 105-A03:2019; each step also carries a tolerance.

Two things follow. First, the steps roughly double — the scale is closer to how the eye perceives difference than a linear scale would be. Second, the staining scale is much wider: a staining grade 4 is a larger colour difference than a colour change grade 3. Comparing "4 for change" with "4 for staining" as if they were the same severity is a misunderstanding of the scales.

The viewing set-up is part of the method

Both standards describe the same viewing arrangement:

In practice, "equivalent source" means a light cabinet with a daylight simulator. A desk under ceiling fluorescents meets none of these points.

Masks: the clause most labs miss

ISO 105-A03:2019 adds a requirement that matters more than it looks: samples must be masked with material identical in colour to the sleeve used on the grey scale. If the scale sits in a neutral grey sleeve, mask the samples in the same grey; if it sits in a black sleeve, use the same black.

The colour around a sample changes how its contrast is perceived. A specimen viewed against a white bench and a grey scale viewed inside a grey sleeve are being judged by two different eyes.

Five-step or nine-step

With a 5-step scale, you give the grade whose contrast equals the specimen's; if the specimen sits nearer the imaginary midpoint between two pairs, you give the half grade. With a 9-step scale, you give the grade whose contrast is nearest.

In both cases, a 5 is given only when there is no perceived difference. "Almost nothing" is 4–5.

Habits that move the grade

When to measure instead

ISO 105-A04 and ISO 105-A05 provide instrumental methods for staining and colour change. They remove assessor variation but bring their own requirements for instrument and procedure. Many labs use them to settle disputes and keep visual assessment for routine work. Either way, the report should say which was used.

Checklist

  1. Right scale: A02 for colour change, A03 for staining?
  2. Light cabinet, ≥ 600 lx, about 45° / 0°, grey N5 surround?
  3. Samples masked in the same colour as the scale's sleeve?
  4. Specimen compared with the correct original?
  5. Same-grade pairs cross-checked at the end?