A fluorescent UV weathering test exposes specimens to ultraviolet light from fluorescent lamps, heat and moisture in repeating cycles, to compare how materials fade, chalk, crack or lose strength. ISO 4892-3 is the ISO method for plastics; ASTM G154 is the ASTM practice for operating the same type of apparatus, and ASTM D4329 applies it to plastics. Textiles, coatings and paper use the same apparatus under their own product standards.
Four settings define the test: lamp type, irradiance, black-panel temperature and the wetting cycle. A report that gives only the number of hours says almost nothing.
Three lamp types, three different tests
| Lamp | ISO 4892-3 description | Typical use |
|---|---|---|
| UVA-340 (type 1A) | Peak at 343 nm; less than 1 % of output below 300 nm. Simulates global solar UV from 300 to 340 nm. | The usual choice for outdoor weathering comparisons |
| UVA-351 (type 1B) | Peak at 353 nm; less than 1 % below 310 nm. Simulates the UV part of sunlight through window glass. | Indoor and behind-glass exposure |
| UVB-313 (type 2) | Emits significantly below 300 nm, shorter than terrestrial sunlight. | Fast screening; can cause degradation that does not happen outdoors |
Because the spectra differ, results from different lamp types should not be compared. ISO 4892-3 also warns against comparing results from different apparatus types or lamps unless a correlation has been established.
The ISO 4892-3:2013 exposure cycles
Table 4 of ISO 4892-3:2013 lists these cycles. Irradiance is measured at 340 nm (310 nm for UVB-313), and the black-panel temperature is the set point during the light step.
| Cycle | Steps | Lamp | Irradiance | Black-panel |
|---|---|---|---|---|
| 1 | 8 h dry UV / 4 h condensation, lamps off | UVA-340 | 0.76 W/(m²·nm) | 60 °C / 50 °C |
| 2 | 8 h dry UV / 0.25 h water spray / 3.75 h condensation | UVA-340 | 0.76 W/(m²·nm) | 50 °C / not controlled / 50 °C |
| 3 | 5 h dry UV / 1 h water spray | UVA-340 | 0.83 W/(m²·nm) | 50 °C / not controlled |
| 4 | 5 h dry UV / 1 h water spray | UVA-340 | 0.83 W/(m²·nm) | 70 °C / not controlled |
| 5 | 24 h dry UV, no moisture | UVA-351 | 0.76 W/(m²·nm) | 50 °C |
| 6 | 8 h dry UV / 4 h condensation | UVB-313 | 0.48 W/(m²·nm) at 310 nm | 70 °C / 50 °C |
The black-panel tolerance is ±3 °C around the set point at equilibrium. During water spray the black-panel temperature is not controlled but should not exceed 30 °C, and the spray water temperature can affect the result. Higher irradiance may be used if all parties agree, at the cost of shorter lamp life.
ISO 4892-3 has been revised since 2013, and ASTM G154 has its own cycle table. Before setting up a programme, check the cycle in the current edition of the standard your specification names.
Hours are not years
There is no fixed conversion from test hours to outdoor exposure. Outdoor weathering depends on climate, orientation, season and pollution, and the laboratory cycle concentrates only some of these factors. ISO 4892-3 therefore recommends exposing a control material of known performance alongside the test specimens, and comparing the specimens with it rather than with a calendar.
For the same reason, a pass after a certain number of hours is only meaningful against an agreed specification, cycle and lamp type.
Where results go wrong
- Uncontrolled irradiance. Fluorescent lamps lose output as they age. Without closed-loop irradiance control and regular calibration, the same programme delivers less UV each month. Follow the lamp maintenance and rotation instructions.
- Specimens at the edge of the exposure area. Irradiance is not uniform. ISO 4892-1 requires specimen repositioning when irradiance in the exposure area falls below 90 % of the peak.
- Empty positions. Fill all spaces in the exposure area, with blank panels if necessary, so that airflow and temperature stay uniform.
- Stressed specimens. Mount specimens without applied stress, unless stress is part of the test.
- Comparing with the shielded area only. A covered part of the specimen is useful for watching progress, but reported results must be based on comparison with file specimens stored in the dark.
- Too few replicates. ISO 4892-3 recommends at least three replicates of each material for statistical evaluation.
What the report should contain
Standard and edition · method and cycle number · lamp type · irradiance and wavelength of control · black-panel temperature · wetting method (condensation or spray) · total exposure time or radiant exposure · control material · how property change was measured
Choosing an instrument
The apparatus has to match the cycle you need. A cycle with condensation or water spray needs a chamber with those stages; a dry cycle such as cycle 5 does not.
The UV weathering tester uses eight fluorescent lamps (four per side), closed-loop irradiance control, condensation and water spray. The benchtop UV tester has one bank of four lamps with the same control and moisture stages. The TL-K035M ECO is for dry UVA exposure only, with no condensation or spray. In each case the lamp type (UVA-340, UVA-351 or UVB-313) is specified with the method at order.
Frequently asked questions
How many hours of UV testing equal one year outdoors?
There is no valid general conversion. Compare against a control material of known outdoor performance exposed in the same test.
Is fluorescent UV the same as a xenon arc test?
No. Xenon arc lamps (ISO 4892-2, ISO 105-B02) reproduce the full solar spectrum including visible light, so they are used for colour fastness to light. Fluorescent UV concentrates on the short-wave UV region. The two do not replace each other.
Can I use UVB-313 lamps to finish faster?
Only if the specification calls for them. UVB-313 emits below the solar cut-off and can produce degradation that does not occur outdoors.