Yarn evenness describes how much the linear density of a sliver, roving or yarn varies along its length. On a capacitive tester the strand runs between the plates of a capacitor; changes in mass change the capacitance, and the instrument turns that signal into unevenness values and counts of faults. ISO 16549 and ASTM D1425 both describe this capacitance method.
CV% and U%
- CV% (coefficient of variation unevenness) is the standard deviation of linear density as a percentage of the mean.
- U% (mean deviation unevenness) is the average absolute deviation as a percentage of the mean.
ISO 16549 prefers CV% and describes U% as outdated, though still allowed. When the mass distribution is close to normal, CV% is about 1.25 times U%. The standard warns that the ratio can differ considerably when the distribution is not normal, so use it with care, for example when converting an old quality table from U% to CV%.
Faults: thin places, thick places and neps
ISO 16549:2021 defines the default counting thresholds and reports counts per 1 000 m:
| Fault | Threshold in ISO 16549 |
|---|---|
| Thin place | Linear density less than 50 % of the mean (−50 %) |
| Thick place | At least 50 % above the mean (+50 %) |
| Nep | Above +200 % |
Any other threshold must be stated in the report. A count at −40 % or +35 % is not the same number as a count at −50 % or +50 %, and the two cannot be compared.
Settings that change the result
| Setting | What the standards say |
|---|---|
| Test speed | ISO 16549 recommends 25 m/min for slivers, 50 m/min for rovings and 400 m/min for yarns; other speeds by agreement, stated in the report. ASTM D1425-96 lists 4 to 400 m/min. |
| Length tested | CV% rises as the tested length increases; ISO 16549 notes that above about 100 m the increase becomes small. ASTM D1425 reads the value after at least 40 m. |
| Conditioning | Standard atmosphere of ISO 139, uniform moisture along the strand. ISO 16549: 24 h is usually enough for unsupported samples, 48 h for packages. |
| Low-twist filament | The yarn flattens in the sensor and reads too uneven; twist must be inserted during testing. |
| Rounding | ISO 16549 rounds results to 0.1 %. |
ASTM D1425 also notes that different instrument models can give different results for the same yarn, because the measuring length and integrator differ. Report the instrument and settings with every value.
Look beyond the single number
One CV% value summarises a complicated property. ASTM D1425 recommends supplementing it with the diagram and the spectrogram. A spectrogram shows periodic variation, which usually points to a drafting or roller fault at a definite wavelength. A rising CV% with a clean spectrogram points to fibre or random drafting problems instead.
Capacitive or optical?
A capacitive sensor measures mass. An optical sensor measures diameter, which also responds to hairiness and yarn structure. Both are useful, but their CV% values are not interchangeable. Compare like with like, and state which principle was used.
What the report should contain
Standard and edition · material and linear density · instrument and measuring principle · test speed · length tested per package · number of packages · CV% (or U%) per package and mean · fault thresholds and counts per 1 000 m · conditioning
Choosing an instrument
The yarn evenness tester uses a capacitive sensor and reports U%, CV%, thin places, thick places and neps for yarns, rovings and slivers. The OPT01 measures yarn diameter optically at 0.5 mm intervals at 20–300 m/min and adds hairiness and a spectrogram; its evenness figures are diameter-based.
Frequently asked questions
Can I convert U% to CV%?
Approximately, by multiplying by 1.25, but only when the mass distribution is close to normal. For a contract value, measure CV% directly.
Why is my CV% higher than the spinning mill's?
Check the test length, speed, conditioning and instrument model first. Different settings give different numbers for the same yarn.
Are thin, thick and nep counts comparable between laboratories?
Only when the same thresholds are used and reported. The default ISO 16549 thresholds are −50 %, +50 % and +200 %.