COLOUR MEASUREMENT

Portable spectrophotometer basics: CIELAB, ΔE*ab versus CMC, measuring geometry and SCI/SCE

November 24, 2026 · 6 min read

A spectrophotometer measures how much light a sample reflects at each wavelength. From that curve it calculates tristimulus values X, Y and Z for a chosen illuminant and observer, and from those the CIELAB coordinates L* (lightness), a* (red–green) and b* (yellow–blue). Everything else, including the ΔE number on the screen, is calculated from these values.

Illuminant and observer come first

The same sample gives different L*a*b* values under different illuminant/observer combinations, because the white-point values Xn, Yn and Zn change. ISO 105-J03 names D65/10° as the preferred combination for colour difference calculation and lists five others, including D65/2°, C and A. Always report which one you used.

ΔE*ab: simple, but not uniform

ΔE*ab is the straight-line distance between two points in CIELAB: the square root of ΔL*² + Δa*² + Δb*². It is easy to compute, but CIELAB is not visually uniform. A ΔE*ab of the same size can look like a small difference in one part of colour space and a large one in another. That makes a single ΔE*ab pass/fail limit unreliable across a range of shades.

CMC(l:c): the textile tolerance formula

ISO 105-J03 modifies CIELAB with the CMC equation to make colour space more visually uniform. The result, ΔECMC, defines an ellipsoid around the reference colour, with semi-axes in the lightness, chroma and hue directions that are calculated from the reference's own L*, C*ab and hab. This lets one agreed tolerance be used whatever the reference colour.

What about ΔE00?

CIEDE2000 (ΔE*00) is a later colour difference formula, and both of our portable instruments calculate it alongside ΔE*ab and CMC. The important rule is the same for every formula: a tolerance written for one formula cannot be applied to another. If the buyer's specification says CMC(2:1) 1.0, a ΔE*00 or ΔE*ab value of 1.0 is a different requirement.

Measuring geometry

Geometry is written as illumination angle / viewing angle. A 45/0 instrument lights the sample at 45° and views it from directly above; d/8 (or D/8) instruments light the sample diffusely with an integrating sphere and view it at 8°. Diffuse and 45/0 instruments generally do not give equivalent results, so data from the two types should not be mixed in one specification.

Sphere instruments can report SCI (specular component included) or SCE (specular component excluded). SCI describes the colour of the material regardless of surface gloss; SCE is closer to what the eye sees when gloss changes the appearance. State which mode you used, and do not compare SCI readings with SCE readings.

Repeatability is not accuracy

Instrument specifications usually quote repeatability (for example the standard deviation of repeated ΔE*ab readings on a white tile) and inter-instrument agreement (the difference between two units of the same model on a set of reference tiles). Repeatability shows how stable one unit is; it says nothing about absolute colour accuracy. For shade approval between a mill and a buyer, inter-instrument agreement matters more.

What the report should contain

ISO 105-J03 asks for the instrument and its CIE geometry, the ΔECMC values, the l and c values, the illuminant and observer, the tolerance used for the pass/fail decision, and, where needed, the component differences and the CIELAB values of reference and sample.

Choosing between the LS176 and the DS-620

LS176DS-620
GeometryD/8, SCId/8°, SCI and SCE
Illuminant / observerD65 / 10°D65, TL84, A, CWF and others; 2° and 10°
Aperture8 mmØ11, Ø6 and Ø3 mm
Colour differenceΔE*ab, ΔE*uv, ΔE*94, ΔE*00, CMCColour difference, fastness, whiteness, yellowness, metamerism and colour strength assessments
Inter-instrument agreementΔE*ab ≤ 0.4ΔE*ab ≤ 0.30

The LS176 suits routine D65/10° quality control with CMC or ΔE*00 tolerances. The DS-620 adds SCE, several illuminants for metamerism checks and small apertures for narrow samples. Instrumental results should still be confirmed visually in a light cabinet; see our note on metamerism and light source selection.

Frequently asked questions

Is ΔE*ab 1.0 the same as CMC(2:1) 1.0?

No. They are different formulas, and the same pair of samples gives different numbers.

Why is l = 2 in CMC(2:1)?

ISO 105-J03 fixes the ratio of the three semi-axes so that results agree best with visual assessment of typical textiles; l is usually 2.0 and c is always 1.0.

Can I compare readings from a 45/0 instrument with a d/8 instrument?

Not reliably. Diffuse and 45/0 geometries generally give different results; use the same geometry, mode and aperture for reference and production.