A light booth is only useful if everyone looks under the same light. Lamps age, diffusers get dirty, the supply voltage varies: a cabinet that was compliant on delivery can drift without anyone noticing. The only protection is to measure and record its conditions at regular intervals.
What a visual assessment method specifies
Turkish standard TS 7166 (1989), a method for visual assessment of small colour differences between opaque objects including textiles, gives concrete values for the viewing conditions:
| Parameter | Requirement (TS 7166:1989) |
|---|---|
| Daylight source | Sources reproducing D65 or D75 |
| Colour rendering index | At least 92 for critical matching; colour temperature must first be within 100 K of the target |
| Illuminance at the centre (medium-lightness samples) | 1,080–1,340 lx for critical assessment; 810–1,880 lx for general assessment |
| Very light / very dark samples | Down to 540 lx for very light; up to 2,150 lx for very dark |
| Uniformity | Illuminance constant within ±20 % across the viewing area |
| Surround and background | Surround Munsell grey N6 to N7; background similar to the standard (critical, chroma ≤ 0.2) or N5 to N7 (general, chroma ≤ 0.3) |
| Surround gloss | No more than 15 on the 60° gloss scale |
For glossy, highly saturated samples the standard asks for a black matt material, such as velvet, on the wall opposite the observer. Viewing is at about 45°, avoiding the specular reflection of the source.
Maintenance every 100 hours
The same standard requires the lamps and equipment to be maintained, with these points checked every 100 operating hours:
- all lamps are working;
- parts are cleaned;
- where required, the lamp voltage is measured and adjusted;
- the illuminance level is measured and recorded.
The last point matters most: without a record there is no way to show a customer that the cabinet was within limits on the day of approval.
A simple verification routine
- Illuminance: measure at the centre of the viewing plane with a calibrated lux meter, for each source, after warm-up.
- Uniformity: also measure at the corners of the area in use and check the ±20 % band.
- Lamp hours: log the hours for each source; a source used every day ages differently from one used only for checks.
- Cleanliness and surround: clean the diffusers and keep the walls grey and matt, with no coloured objects inside.
- Log: date, source, measured values, operator. A value outside the limits means a lamp or setting to fix before the next approval.
Illuminance says nothing about spectral distribution. Where matching is critical, or two sites compare results, a spectral measurement of the source is the only way to confirm that it still reproduces the intended daylight.
ISO 3664 is not a textile standard
ISO 3664 is prepared by the photography and graphic technology committees and covers viewing of prints and reproductions. A cabinet labelled ISO 3664 is not automatically right for textile assessment: see our guide Buying a light booth.
Our instruments
PRO models of the light assessment cabinet log operating hours per source, give a lamp replacement warning and track warm-up time; the interior has a matt, anti-reflective finish. D65 sources are replaced with the D65 LED tube lamp. To choose which sources to check, see metamerism in shade approval.
Frequently asked questions
How often should a light booth be checked?
TS 7166 says every 100 operating hours. Follow your customer's or quality system's requirement too if it is stricter.
Is a lux meter enough?
For illuminance and uniformity, yes. To confirm the spectral distribution of the source you need a spectral measurement.
Should a lamp that still works be replaced?
Yes, if illuminance or uniformity fall outside the limits, or the manufacturer flags end of life: a lamp that lights up is not necessarily a compliant lamp.