TEST STANDARD

ISO 105-X11 hot pressing: the 2026 edition changed the temperatures — check your procedures

October 28, 2026 · 5 min read

For three decades, ironing fastness under ISO 105-X11 meant one of three numbers: 110 °C, 150 °C or 200 °C. Those numbers are on worksheets, in buyer manuals and in the presets of many hot-pressing testers.

In June 2026 a new edition, ISO 105-X11:2026, was published, replacing the 1994 edition. Its published summary lists 120 °C, 160 °C and 210 °C as the three commonly used pressing temperatures. If your procedures still say 110/150/200, they now describe the withdrawn edition.

What the test is for

ISO 105-X11 determines the resistance of the colour of textiles to ironing and to processing on hot cylinders. It is relevant wherever colour can change or migrate under heat during pressing — in the garment factory, at the dry cleaner or at home. The standard offers three procedures, chosen according to the end use:

ProcedureWhat is pressed
Dry pressingA dry specimen alone
Damp pressingA dry specimen covered by a wet cotton adjacent fabric
Wet pressingA wet specimen covered by a wet cotton adjacent fabric

Colour change of the specimen is graded with the ISO 105-A02 grey scale and staining of the cotton adjacent fabric with the ISO 105-A03 grey scale.

What the 1994 edition specified

The details of the 2026 edition should be read from the new text itself. For reference, the withdrawn 1994 edition specified:

Both the 1994 summary and the 2026 summary say that for blends the temperature should suit the fibre with the lowest heat resistance, and that other temperatures may be used if they are reported. Before you change anything else, obtain the 2026 text and compare it clause by clause with your procedure — especially pressure, time and the assessment sequence.

What changes in the lab

Worksheets and presets

Every preset, worksheet and LIMS entry that says 110, 150 or 200 °C needs a review. A tester that has been running "ISO 105-X11, 150 °C" for years will keep doing exactly that unless someone changes it.

Reports

A report must name the edition. "ISO 105-X11" with no year, at 150 °C, is now ambiguous: it was correct under 1994 and is a non-standard temperature under 2026 — permitted, but it must be stated. Write "ISO 105-X11:2026, dry pressing, 160 °C" or, during a transition, "ISO 105-X11:1994, 150 °C, as specified by customer".

Customer specifications

Many buyer manuals quote the old temperatures. Do not silently switch: ask the customer which edition and temperature they want, and record the answer. On heat-sensitive dyeings, ten degrees can be the difference between two grades.

Equipment range and calibration

Check that your tester covers the new top temperature with margin and that its temperature has been verified at the new set points, not only at the old ones. A calibration certificate with points at 110, 150 and 200 °C does not show how the plate behaves at 210 °C.

Bench errors that the new edition does not fix

Measuring the wrong temperature

The display shows the controller's sensor, not the plate surface where the specimen sits. Verify the surface temperature with a surface probe or temperature-indicating strips at each set point used.

Losing the pressure

The specified pressure depends on the relationship between the top plate's mass and the padding area. Thick specimens need a larger specimen area or a template, so that the load is not carried by a few high points.

Wet means a defined amount of water

"Wet" in damp and wet pressing is not "dripping". Squeeze or extract to the specified water content and weigh it. Excess water cools the plate and changes the result.

Grading only once

Some heat-induced shade changes partly recover as the fabric cools and regains moisture. That is why the procedure grades colour change both immediately and after conditioning. A single grade taken straight off the press tells only half the story.

Hand irons

The 1994 edition allowed a household iron when no heating device was available, but warned that its uneven surface temperature limits accuracy and reproducibility, and required the report to say so. If a hand iron was used, write it on the report.

AATCC 133 is a separate method

AATCC 133 also covers colour fastness to heat by hot pressing. It is a different document with its own conditions; an ISO 105-X11 result does not satisfy an AATCC 133 requirement, or the reverse.

Checklist

  1. Does the procedure cite ISO 105-X11:2026, and has it been compared with the new text?
  2. Are the temperatures updated, or the customer's choice of edition recorded?
  3. Has the tester been verified at the new set points?
  4. Was water content in damp/wet pressing weighed, not estimated?
  5. Was colour change graded immediately and after conditioning?

TLI-TexLabInstruments manufactures its iron fastness tester, with a 20–220 °C plate range, in its own workshop in Istanbul.

Cenk YILMAZ
TLI-TexLabInstruments